Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | When an evaluation is carried out using a non-standard protocol, no inference can be made on accuracy based on a standard protocol. |
Fassaert LMM, Plate JDJ, Westerink J, Immink RV, de Borst GJ. The ClearSight System for Postoperative Arterial Blood Pressure Monitoring After Carotid Endarterectomy: A Validation Study. Am J Hypertens. 2022 Feb 1;35(2):164-172. doi: 10.1093/ajh/hpab140. PMID: 34505631. Available from: PMC8807161.
81060-2:2009 - Fail Postoperative subjects (n=48), IA validation (Note: Fail for SBP and DBP, even after corrections. Pass for MAP.)
Helmer P, Helf D, Sammeth M, Winkler B, Hottenrott S, Meybohm P, Kranke P. The Use of Non-Invasive Continuous Blood Pressure Measuring (ClearSight®) during Central Neuraxial Anaesthesia for Caesarean Section-A Retrospective Validation Study. J Clin Med. 2022 Aug 2;11(15):4498. doi: 10.3390/jcm11154498. PMID: 35956113. Available from: PMC9369920.
Ad Hoc protocol n=31
Unreferenced "AAMI" partial criteria were "considered acceptable".
Kho E, van der Ster BJP, van der Ven WH, Vlaar APJ, Immink RV, Veelo DP. Clinical agreement of a novel algorithm to estimate radial artery blood pressure from the non-invasive finger blood pressure. J Clin Anesth. Epub: 2022 Sep 26;83:110976 p. doi: 10.1016/j.jclinane.2022.110976. PMID: 36174389. Available from: www.sciencedirect.com.
Ad Hoc protocol n=33
Not independent; Failed Criterion 1 of ISO 81060-2:2019; Error grid criteria used.