Datascope Accutorr Plus
Device Name:
Datascope Accutorr Plus
Manufacturer:

Brand: SOMA Technology Inc., 166 Highland Park Drive, Bloomfield, CT 06002, UNITED STATES.

OBL: Mindray North America, 800 Macarther Blvd, Mahwah, NJ 07430, UNITED STATES.

OEM: Datascope Corp. (Maquet/Getinge), 1300 MacArthur Blvd, Mahwah, NJ 07430, UNITED STATES.

Measuring functions:
Blood pressure, SpO2, Temperature
Primary Client Use:
Intended for patient monitoring
Measurement Site:
BP: Upper Arm; SpO2: Finger
Measurement Occurrence:
BP & SpO2: Single and intermittent measurements
Availability:
Available Currently
Device Manual:
Device Specifications:
Description:
The Datascope Accutorr Plus is a patient monitor. Its blood pressure measurement technology has been proven to be accurate, with a 1-star Medaval rating and information regarding the accuracy of its oxygen saturation measurement technology does not appear to be available. Blood pressure measurements are taken from the upper arm and oxygen saturation measurements are taken from the finger. It is intended for bedside patient monitoring.
Assessment:
The technology used in the Datascope Accutorr Plus, to measure blood pressure, has passed in four clinical validation studies, between general and young populations, according to recognised standard protocols, as published in peer-reviewed publications. There appears to be no peer-reviewed clinical validation information available on the technology used in the Datascope Accutorr Plus to measure peripheral oxygen saturation.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval Recommendation Older clinical validation; older protocol; multiple clinical validations
BP BIHS (UK and IRL) Professional use for adults and children Published evidence
BP ESH (Europe) Professional use for adults and children Published evidence
BP Stride BP Office and hospital use for adults Published evidence
BP FDA (US) 510(k) 510(k) K970061,, K983575
Device Family:
Datascope Accutorr VB, Datascope Accutorr PlusB
Legend: B BIHS Derivative
Validation Publications:

Anwar YA, Tendler BE, McCabe EJ, Mansoor GA, White WB. Evaluation of the Datascope Accutorr Plus according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit. 1997 Apr;2(2):105-110. PMID: 10234101.

SP10:1992 - Pass General population

White WB, Herbst T, Thavarajah S, Giacco S. Clinical evaluation of the Trimline blood pressure cuffs with the Accutorr Plus Monitor. Blood Press Monit. 2003 Jun;8(3):137-40. PMID: 12900592.

SP10:2002 - Pass General population

Trimline blood pressure cuffs used

Wong SN, Tz Sung RY, Leung LC. Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children. Blood Press Monit. 2006 Oct;11(5):281-91. PMID: 16932037.

ESH-IP:2002 - Pass Children

Khawaja RA, Qureshi R, Mansure AH, Yahya ME. Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines. J Saudi Heart Assoc. 2010 Jan;22(1):1-5. Epub: 2010 Mar 10. doi: 10.1016/j.jsha.2010.03.001. PMID: 24936124. Available from: PMC4051388.

BHS:1993 - Pass (A/A) General population

SP10:1996 - Pass General population

Relevant Publications:

Adler C, Ellert U, Neuhauser HK. Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values. Blood Press Monit. 2014 Apr;19(2):109-17. doi: 10.1097/MBP.0000000000000037. PMID: 24583967.

Measurements from the Datascope Accutorr Plus are compared to those from the Omron HEM-705CP II, though not in a formal validation.

Sarganas G, Kuhnert R, Gohlisch C, van der Giet M, Neuhauser H. Comparison of two blood pressure oscillometric devices: Datascope Accutorr Plus and Mobil-O-Graph PWA and conversion of blood pressure values from one device to the other. Blood Press Monit. 2020 Feb;25(1):42-49. Epub: 2019 Oct 15. doi: 10.1097/MBP.0000000000000409. PMID: 31633519.

Measurements from the Datascope Accutorr Plus are compared to those from the IEM Mobil-O-Graph PWA, though not in a formal validation.