Citizen CH-452
Device Model:
CH-452
Manufacturer:
Citizen Systems Japan Co. Ltd., 6-1-12 Tanashi-cho, Nishi-Tokyo-Shi, Tokyo 188-8511, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Device Manual:
Description:
The Citizen CH-452 is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use.
Assessment:
The technology used in the Citizen CH-452, to measure blood pressure, has been compared to a clinically validated device, but not according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | ESH (Europe) | Self-measurement | Published evidence |
BP | Japanese Society of Hypertension | Self-measurement (Latest: 2016-2018, 2021-2022) | Unpublished equivalence to the Citizen CH-432B, by questionnaire |
Device Family:
Citizen CH-432BD,J, Citizen CH-452D,J, Citizen CH-432J, Citizen CH-453FJ, Citizen CH-550J, Citizen CHU302J
Legend: D DET Equivalence, J JHS Implied Equivalence
Legend: D DET Equivalence, J JHS Implied Equivalence
Validation Publications for Equivalent Devices:
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Citizen CH-432B
Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C. Blood Press Monit. 2008 Feb;13(1):55-62. doi: 10.1097/MBP.0b013e3282cb599a. PMID: 18199925. Erratum in: Blood Press Monit. 2008 Aug;13(4):243.
ESH-IP:2002 - Pass General population