OBL: BpTru (Appears to be Suspended), Unit 1, 1850 Hartley Avenue Coquitlam, BC Canada V3K 7, CANADA.
OEM: VSM MedTech Ltd. (Appears to be Suspended), 9 Burbidge Street, Coquitlam, British Columbia V3K 7B2, CANADA.
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | BIHS (UK and IRL) | Previous recommendation, now archived (Archived) | Manufacturer declaration of equivalence to the BpTRU BPM-100 |
BP | ESH (Europe) | Professional use for adults and children | The BIHS recommendation, despite the absence, at the time, of the scientific evidence required by the same authors' criteria. |
BP | FDA (US) | 510(k) | 510(k) K023055 |
Padwal RS, Majumdar SR. Comparability of two commonly used automated office blood pressure devices in the severely obese. Blood Press Monit. 2016 Oct;21(5):313-5. doi: 10.1097/MBP.0000000000000196. PMID: 27100401 .
Measurements from the BpTru BPM-200 are compared to those from the Microlife Watch BP Office, though not in a formal validation.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
BpTru BPM-100
Mattu GS, Perry TL Jr, Wright JM. Comparison of the oscillometric blood pressure monitor (BPM-100(Beta)) with the auscultatory mercury sphygmomanometer. Blood Press Monit. 2001 Jun;6(3):153-9. PMID: 11518839.
SP10:1996 - Pass General population
BHS:1993 - Pass (A/A) General population
Wright JM, Mattu GS, Perry Jr TL, Gelferc ME, Strange KD, Zorn A, Chen Y. Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Press Monit. 2001 Jun;6(3):161-5. PMID: 11518840.
SP10:1996 - Pass General population
BHS:1993 - Pass (A/A) General population
Mattu GS, Heran BS, Wright JM. Comparison of the automated non-invasive oscillometric blood pressure monitor (BpTRU) with the auscultatory mercury sphygmomanometer in a paediatric population. Blood Press Monit. 2004 Feb;9(1):39-45. PMID: 15021077.
SP10:1992 - Pass Children, aged 3 to 18 years (n=162/36)
BHS:1993 - Pass (B/B) Children, aged 3 to 18 years (n=162/36)
Mattu GS, Heran BS, Wright JM. Overall accuracy of the BpTRU--an automated electronic blood pressure device. Blood Press Monit. 2004 Feb;9(1):47-52. PMID: 15021078.
BHS:1993 - Pass (A/A) General population
SP10:2002 - Pass General population
Tran K, Potts J, Purkiss S, Robertson J, Khan N, Padwal R, Chan WS. Validation of an Automated Office Blood Pressure Machine in Pregnant Women According to the AAMI 2013/ISO Protocol. Hypertension. 2018 Dec;72(6):e91-e94. Epub: 2018 Oct 22. doi: 10.1161/HYPERTENSIONAHA.118.12085. PMID: 30571239. Research Letter.
81060-2:2013 - Fail Normotensive pregnancy (n=15), Hypertensive pregnancy (n=15), Pre-eclampsia (n=15) (Note: Overall results. Results for subgroups are provided for bands only.)