Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | BIHS (UK and IRL) | Previous recommendation, now archived (Archived) | Manufacturer declaration of equivalence to the Omron MX3 Plus (HEM-742-E) |
BP | ESH (Europe) | Self-measurement | The BIHS recommendation, despite the absence, at the time, of the scientific evidence required by the same authors' criteria. |
Ware A, Stevens R, Selwood M, Fleming S. Accuracy of blood pressure monitors available in high street pharmacies. Blood Press Monit. 2016 Feb;21(1):59-61. doi: 10.1097/MBP.0000000000000151. PMID: 26427055.
Measurements from the Boots 5690447 are compared to those from the Omron PA-350 (BPM tester), though not in a formal validation.
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Omron MX3 Plus (HEM-742-E)
Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol. Blood Press Monit. 2005 Jun;10(3):165-8. PMID: 15923819.
ESH-IP:2002 - Pass General population
Christofaro DG, Casonatto J, Polito MD, Cardoso JR, Fernandes R, Guariglia DA, Gerage AM, de Oliveira AR. Evaluation of the Omron MX3 Plus monitor for blood pressure measurement in adolescents. Eur J Pediatr. 2009 Nov;168(11):1349-54. Epub: 2009 Feb 17. doi: 10.1007/s00431-009-0936-x. PMID: 19221789.
BHS:1993 - Pass (A/A) General population
SP10:1996 - Pass General population