Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
BG | MDR Criteria | Self-measurement and professional use | Published evidence |
Legend: Z Almost Certain
Pleus S, Baumstark A, Jendrike N, Mende J, Link M, Zschornack E, Haug C, Freckmann G. System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015. BMJ Open Diabetes Res Care. 2020 Jan;8(1):e001067. doi: 10.1136/bmjdrc-2019-001067. PMID: 31958308. Available from: drc.bmj.com.
15197:2013 - Pass General population (Note: Modified protocol: One reagent system lot.)
Freckmann G, Baumstark A, Jendrike N, Mende J, Schauer S, Link M, Pleus S, Haug C. Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems. J Diabetes Sci Technol. 2022 Sep;16(5):1076-1088. Epub: 2020 Aug 19. doi: 10.1177/1932296820948873. PMID: 32814455. Available from: PMC9445332.
15197:2013 - Pass General population
15197:2013 - Pass General population
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Beurer GL 50 evo mg/dl
Pleus S, Baumstark A, Jendrike N, Mende J, Link M, Zschornack E, Haug C, Freckmann G. System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015. BMJ Open Diabetes Res Care. 2020 Jan;8(1):e001067. doi: 10.1136/bmjdrc-2019-001067. PMID: 31958308. Available from: drc.bmj.com.
15197:2013 - Pass General population (Note: Modified protocol: One reagent system lot.)
Kenning M, Puchert A, Salzsieder E. System Accuracy Evaluation of Blood Glucose Monitoring Systems: Comparison of Four Devices for Personal Use Pursuant to ISO and ADA Guidelines. J Diabetes Sci Technol. 2021 Sep;15(5):1202-1203. Epub: 2021 May 12. doi: 10.1177/19322968211011185. PMID: 33978521. Available from: PMC8442185.
15197:2013 - Pass General population
FDA:2020H - Pass American Diabetes Accociation (ADA) protocol claimed but ADA reference refers to FDA 2020 and ISO 15197:2013
Beurer GL 50 mg/dl
Hasslacher C, Kulozik F, Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting. J Diabetes Sci Technol. 2014 May;8(3):466-72. Epub: 2014 Feb 18. doi: 10.1177/1932296814522804. PMID: 24876607. Available from: PMC4455429.
15197:2013 - Fail General population