Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
Legend: Z Almost Certain
Kenning M, Puchert A, Salzsieder E. System Accuracy Evaluation of Blood Glucose Monitoring Systems: Comparison of Four Devices for Personal Use Pursuant to ISO and ADA Guidelines. J Diabetes Sci Technol. 2021 Sep;15(5):1202-1203. Epub: 2021 May 12. doi: 10.1177/19322968211011185. PMID: 33978521. Available from: PMC8442185.
15197:2013 - Pass General population
FDA:2020H - Pass American Diabetes Accociation (ADA) protocol claimed but ADA reference refers to FDA 2020 and ISO 15197:2013
Kenning M, Puchert A, Salzsieder E. System Accuracy Evaluation of a Blood Glucose Monitoring System. J Diabetes Sci Technol. 2022 Mar;16(2):574-575. Epub: 2021 Nov 15. doi: 10.1177/19322968211057534. PMID: 34781779. Available from: PMC8861790.
15197:2013 - Pass General population
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Beurer GL 48 mg/dl
Kenning M, Puchert A, Salzsieder E. System Accuracy Evaluation of Blood Glucose Monitoring Systems: Comparison of Four Devices for Personal Use Pursuant to ISO and ADA Guidelines. J Diabetes Sci Technol. 2021 Sep;15(5):1202-1203. Epub: 2021 May 12. doi: 10.1177/19322968211011185. PMID: 33978521. Available from: PMC8442185.
15197:2013 - Pass General population
FDA:2020H - Pass American Diabetes Accociation (ADA) protocol claimed but ADA reference refers to FDA 2020 and ISO 15197:2013