OBL: Beurer GmbH, Söflinger Straße 218, 89077 Ulm, GERMANY.
OEM: Nihon Seimitsu Sokki Co. Ltd. (Nissei), 2508-13 Nakago Shibukawa, Gunma 377-0293, JAPAN.
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | Legacy approval | Older clinical validation; older protocol |
BP | DHL (DE) | General use (Previous: 2011) | Published evidence |
BP | ESH (Europe) | Self-measurement | The DHL recommendation |
Consumer Body | Year(s) | Level | Comment | |
BP | Choice (Australia) | Previous | B |
Tholl U, Lüders S, Bramlage P, Dechend R, Eckert S, Mengden T, Nürnberger J, Sanner B, Anlauf M. The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control. Blood Press Monit. 2016 Aug;22(4):197-205. doi: 10.1097/MBP.0000000000000186. PMID: 26998590.
DHL:1999 - Pass General population (Note: 2011)
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Nissei DS-400
Duhig KE, De Greeff A, Van Der Westhuizen A, Baker E, Shennan AH. Validation of the Nissei DS-400 in a low-resource setting. Blood Press Monit. 2009 Jun;14(3):132-5. doi: 10.1097/MBP.0b013e32832ce136. PMID: 19433979.
ESH-IP:2002 - Pass Low resource setting
de Greeff A, Shennan AH. Clinical accuracy of a low cost portable blood pressure device in pregnancy and pre-eclampsia: the Nissei DS-400. Trop Doct. 2015 Jul;45(3):168-73. Epub: 2015 Apr 23. doi: 10.1177/0049475515581542. PMID: 25911652.
BHS:1993 - Pass (A/A) Pregnancy and pre-eclampsia (n=45)