🏠
📞  +353 1 6643768
 
📧  info@medaval.ie
Aspen C5
Device Name:
C5
Manufacturer:

OBL: Aspen Diagnostics Pvt. Ltd., ℅ Rapid Diagnostic Pvt. Ltd., B-82, GT Karnal Road Industrial Area, Ashok Vihar, Delhi 110033, INDIA.

OEM: Rossmax International Ltd., 12F - 189 Kang Chien Road, Taipei 114, TAIWAN ROC.

Measuring functions:
Blood pressure
Primary Client Use:
Intended for professional use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Available Currently
Description:
The Aspen C5 is an automatic blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for professional use.
Assessment:
While the technology used in the Aspen C5, to measure blood pressure, has been declared as being equivalent to that used in a clinically validated device, no evidence has been published to show that the devices have been compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP BIHS (UK and IRL) Self-measurement Manufacturer declaration of equivalence
Device Family:
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Rossmax CF175f (V1)

Zhang L, Kang YY, Zeng WF, Li Y, Wang JG. Validation of the Rossmax CF175 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2015 Apr;20(2):104-7. doi: 10.1097/MBP.0000000000000089. PMID: 25350783.

ESH-IP:2010 - Pass General population