OBL: WEPA Apothekenbedarf GmbH & Co KG (Aponorm), Am Fichtenstrauch 6-10, 56204 Hillscheid, GERMANY.
OEM: Microlife Corporation, 9F - 431 RuiGuang Road, NeiHu, Taipei 11492, TAIWAN ROC.
Accuracy Assessment | Recommendation | Basis | |
BP | Medaval | None | Although equivalence to another device is claimed, this has not been tested to MDR requirements. |
BP | DHL (DE) | General use (Latest: 2019) | Manufacturer declaration of equivalence |
BP | Stride BP | Self-measurement for adults (Preferred) | Equivalence to the Microlife BP B3 AFib (BP 3KT1-3N) claim without published evidence |
Legend: B BIHS Derivative, P Stride Equivalence, Z Almost Certain
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
Microlife BP B3 AFib (BP 3KT1-3N)
Beime B, Bramlage C, Krüger R, Deutsch C, van Mark G, Bramlage P, Botta B. Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2: 2019 protocol. Blood Press Monit. 2021 Aug 1;26(4):299-304. Epub: 2021 Mar 15. doi: 10.1097/MBP.0000000000000530. PMID: 33734119.
81060-2:2018 - Pass General population
Microlife BP B6 Connect (BP 3KV1-5X)
Menti A, Ntousopoulos V, Lempidaki P, Bakalakou KR, Komnianou A, Mariglis D, Ntaskagianni E, Kollias A, Stergiou GS. Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KV1-5X (BP B6 Connect) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018). Blood Press Monit. Epub: 2025 Sep 10. doi: 10.1097/MBP.0000000000000771. PMID: 40931782.
81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population