🏠
📞  +353 1 6643768
 
📧  info@medaval.ie
Andon KD-5965
Device Name:
KD-5965
Device Model:
KP-5965
Manufacturer:
Andon Health Co. Ltd., 3 Jinping Road, Ya'an Street, Nankai District, Tianjin 300190, CHINA.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Device Accessibility:
Voiced instructions and results
Availability:
Discontinued but still available
Device Specifications:
Description:
The Andon KD-5965 is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a certified 4-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The accuracy of the technology used in the Andon KD-5965, to measure blood pressure, is certified by Medaval Ltd. It has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★★★ Certified Accuracy Recent clinical validation; older protocol with certified adherence
BP BIHS (UK and IRL) Self-measurement (Previous) Published evidence
BP ESH (Europe) Self-measurement Published evidence
BP Stride BP Self-measurement and professional use for adults (Unavailable) (Preferred) Published evidence
Device Family:
Validation Publications:

Huang J, Li Z, Li G, Liu Z. Validation of the Andon KD-5965 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2015 Oct;20(5):283-5. Epub: 2015 May 5. doi: 10.1097/MBP.0000000000000129. PMID: 25968093.

ESH-IP:2010 - Pass General population

Medaval Certification Assessment

Certification ImageMedaval Ltd. Accreditation assessment of the blood pressure measurement technology used in the Andon KD-5965 (KP-5965) upper arm monitor, as validated according to the European Society of Hypertension International Protocol revision 2010. Medical Device Assessment. 2016 Aug 5;2016(1608) 5 p. Epub: 2019 Jan 31. Available from: www.medaval.ie.

With proven validity of validation results according to an accepted protocol, the clinical accuracy of the blood pressure measurement technology, used in the Andon KD-5965, is certified by Medaval Ltd.

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Andon KD-5031

Wu N, Zhang X, Wang W, Zhang H. Validation of the Andon KD5031 for clinical use and self-measurement according to the European Society of Hypertension International Protocol. Blood Press Monit. 2016 Oct;21(5):310-2. Epub: 2016 Jul 7. doi: 10.1097/MBP.0000000000000204. PMID: 27392206.

ESH-IP:2010 - Pass General population

Andon KD-5915 (V1)

Huang QF, Wang J, Sheng CS, Zhang NN, Li Y, Wang JG. Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit. 2010 Aug;15(4):232-4. PMID: 20658765.

ESH-IP:2002 - Pass General population

Andon KD-5917

Guo WG, Li BL, He Y, Xue YS, Wang HY, Zheng QS, Xiang DC. Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2014 Aug;19(4):242-5. doi: 10.1097/MBP.0000000000000048. PMID: 24847724.

ESH-IP:2010 - Pass General population

Andon KD-595

Wu N, Zhang X, Wang W, Zhang H. Validation of the Andon KD-595 for clinical use and self-measurement according to the European Society of Hypertension International Protocol. Blood Press Monit. 2016 Apr;21(2):124-7. Epub: 2015 Nov. doi: 10.1097/MBP.0000000000000164. PMID: 26544523.

ESH-IP:2010 - Pass General population

Liu K, Wu D, Bing S, Li LY. Validation of the ANDON KD-595 automated upper-arm blood pressure monitor according to the universal standard. Blood Press Monit. 2024 Apr 1;29(2):103-106. Epub: 2023 Dec 14. doi: 10.1097/MBP.0000000000000686. PMID: 38193382.

81060-2:2019 - Pass General population

Beurer BM 44

Lüders S, Krüger R, Zemmrich C, Forstner K, Sturm CD, Bramlage P. Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002. Blood Press Monit. 2012 Dec;17(6):248-52. doi: 10.1097/MBP.0b013e32835b9e8e. PMID: 23147534.

ESH-IP:2002 General population (Note: ESH-02 not recognised since 01 July 2011)

iHealth Track (KN-550BT)

Van Den Heuvel JFM, Lely AT, Franx A, Bekker MN. Validation of the iHealth Track and Omron HEM-9210T automated blood pressure devices for use in pregnancy. Pregnancy Hypertens. 2019 Jan;15:37-41. Epub: 2018 Nov 2. doi: 10.1016/j.preghy.2018.10.008. PMID: 30825925.

ESH-IP:2010 - Pass Pregnancy (n=33), Pre-eclampsia (n=10)

Shi L, Zhang XY, Sun JY, Chen JF, Ma YF, Zhou KY. Validation of the iHealth Track upper-arm blood pressure monitor KN-550BT in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020). Blood Press Monit. 2024 Feb 1;29(1):41-44. Epub: 2023 Sep 14. doi: 10.1097/MBP.0000000000000678 . PMID: 37702596.

81060-2:2020 - Pass General population