Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
BG | FDA (US) | 510(k) | 510(k) K103544 |
BG | TÜV Rheinland Nederland | General use | Unpublished internal data |
BG | MDR Criteria | Self-measurement | Published evidence |
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.
15197:2003 - Fail General population
15197:2013 - Fail General population
Pfützner A, Mitri M, Musholt PB, Sachsenheimer D, Borchert M, Yap A, Forst T. Clinical assessment of the accuracy of blood glucose measurement devices. Curr Med Res Opin. 2012 Apr;28(4):525-31. Epub: 2012 Mar 21. doi: 10.1185/03007995.2012.673479. PMID: 22435798.
15197:2003 - Pass General population
Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: PMC3876330.
15197:2003 - Pass Using insufficient blood samples (n=31)
Pfützner A, Hengesbach C, Demircik F, Schipper C, Forst T, Musholt PB. Performance of blood glucose meters in compliance with current and future clinical ISO15197 accuracy criteria. Curr Med Res Opin. 2014 Feb;30(2):185-90. Epub: 2013 Nov 6. doi: 10.1185/03007995.2013.858617. PMID: 24156775.
15197:2003 - Pass General population
15197:2013 - Pass General population
Hasslacher C, Kulozik F, Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting. J Diabetes Sci Technol. 2014 May;8(3):466-72. Epub: 2014 Feb 18. doi: 10.1177/1932296814522804. PMID: 24876607. Available from: PMC4455429.
15197:2013 - Pass General population
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
AgaMatrix Wavesense Jazz
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.
15197:2003 - Fail General population
15197:2013 - Fail General population
Grant DA, Dunseath GJ, Bain SC, Luzio S. Postmarket Approval Surveillance of a Low Acquisition Cost Blood Glucose Monitoring System: Assessment of the Accuracy Following ISO 15197:2013. J Diabetes Sci Technol. 2016 Sep;10(5):1195-6. Epub: 2016 May 12. doi: 10.1177/1932296816648991. PMID: 27179012. Available from: PMC5032959.
15197:2013 - Pass General population