Accuracy Assessment | Recommendation | Basis | |
BG | Medaval | ★★★ Recommendation | Recent clinical validation; recent protocol; multiple clinical validations |
BG | TÜV Rheinland Nederland | General use | Unpublished internal data |
BG | Institute of Health Economics (CA) | General use | No evidence provided |
BG | MDR Criteria | Self-measurement | Published evidence |
Consumer Body | Year(s) | Level | Comment | |
BG | Consumer Reports (US) | Latest: 2019-2022 | Excellent | |
BG | Choice (Australia) | Latest: 2019 | A | |
BG | Amazon Best Sellers (US) | Latest: 2019 | A (4.0 / 5) |
Tack C, Pohlmeier H, Behnke T, Schmid V, Grenningloh M, Forst T, Pfützner A. Accuracy evaluation of five blood glucose monitoring systems obtained from the pharmacy: a European multicenter study with 453 subjects. Diabetes Technol Ther. 2012 Apr;14(4):330-7. Epub: 2011 Dec 16. doi: 10.1089/dia.2011.0170. PMID: 22176154.
15197:2003 - Pass General population
15197:2013 - Pass General population
Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: PMC3570840.
15197:2003 - Pass General population
15197:2013 - Pass General population
Pfützner A, Mitri M, Musholt PB, Sachsenheimer D, Borchert M, Yap A, Forst T. Clinical assessment of the accuracy of blood glucose measurement devices. Curr Med Res Opin. 2012 Apr;28(4):525-31. Epub: 2012 Mar 21. doi: 10.1185/03007995.2012.673479. PMID: 22435798.
15197:2003 - Pass General population
Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: PMC3876330.
15197:2003 - Fail Using insufficient blood samples (n=31)
Halldorsdottir S, Warchal-Windham ME, Wallace JF, Pardo S, Parkes JL, Simmons DA. Accuracy evaluation of five blood glucose monitoring systems: the North American comparator trial. J Diabetes Sci Technol. 2013 Sep 1;7(5):1294-304. doi: 10.1177/193229681300700520. PMID: 24124957. Available from: PMC3876374.
Ad Hoc protocol General population
Pfützner A, Hengesbach C, Demircik F, Schipper C, Forst T, Musholt PB. Performance of blood glucose meters in compliance with current and future clinical ISO15197 accuracy criteria. Curr Med Res Opin. 2014 Feb;30(2):185-90. Epub: 2013 Nov 6. doi: 10.1185/03007995.2013.858617. PMID: 24156775.
15197:2003 - Pass General population
15197:2013 - Pass General population
Jendrike N, Baumstark A, Pleus S, Liebing C, Kamecke U, Haug C, Freckmann G. Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. Curr Med Res Opin. 2019 Feb;35(2):301-311. Epub: 2018 Jul 19. doi: 10.1080/03007995.2018.1491832. PMID: 29927663.
15197:2013 - Pass General population
Pleus S, Baumstark A, Jendrike N, Mende J, Link M, Zschornack E, Haug C, Freckmann G. System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015. BMJ Open Diabetes Res Care. 2020 Jan;8(1):e001067. doi: 10.1136/bmjdrc-2019-001067. PMID: 31958308. Available from: drc.bmj.com.
15197:2013 - Pass General population (Note: Modified protocol: One reagent system lot.)
Freckmann G, Baumstark A, Jendrike N, Mende J, Schauer S, Link M, Pleus S, Haug C. Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems. J Diabetes Sci Technol. 2022 Sep;16(5):1076-1088. Epub: 2020 Aug 19. doi: 10.1177/1932296820948873. PMID: 32814455. Available from: PMC9445332.
15197:2013 - Pass General population
15197:2013 - Fail General population
Musholt PB, Schipper C, Thomé N, Ramljak S, Schmidt M, Forst T, Pfützner A. Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices. J Diabetes Sci Technol. 2011 Sep 1;5(5):1167-75. doi: 10.1177/193229681100500520. PMID: 22027312. Available from: PMC3208875.
Performance with haematocrit variation unstable.
Perera NJ, Molyneaux L, Constantino MI, McGill M, Yue DK, Twigg SM, Ross GP. Suboptimal performance of blood glucose meters in an antenatal diabetes clinic. Diabetes Care. 2011 Feb;34(2):335-337. Epub: 2011 Jan 7. doi: 10.2337/dc10-1457. PMID: 21216849. Available from: PMC3024344.
The publication evaluates the performance of the Abbott FreeStyle Lite, though not in a formal validation.
Fournier PA. Analysis of "Accuracy evaluation of five blood glucose monitoring systems: the North American comparator trial". J Diabetes Sci Technol. 2013 Sep 1;7(5):1305-7. doi: 10.1177/193229681300700521. PMID: 24124958. Available from: PMC3876375.
The publication is a letter or correspondence in which the Abbott FreeStyle Freedom Lite is referenced.
Demircik F, Ramljak S, Hermanns I, Pfützner A, Pfützner A. Evaluation of Hematocrit Interference With MyStar Extra and Seven Competitive Devices J Diabetes Sci Technol. 2015 Mar;9(2):262-267. Epub: 2014 Dec 30. doi: 10.1177/1932296814565790. PMID: 25549636. Available from: PMC4604595.
Good accuracy but potential haematocrit influence.