AViTA BPM64 (V1)
Device Model:
BPM64 (Note: 1 of 2)
Manufacturer:
AViTA Corporation, 9F No. 78 Sec 1, Gwang-Fu Road, Sanchong Dist., New Taipei City 24158, TAIWAN ROC.
Measuring functions:
Blood Pressure and PR
Primary Client Use:
Intended for self-measurement and home use
Measurement Site:
Upper Arm
Measurement Occurrence:
Single measurements only
Availability:
Discontinued but still available
Description:
The AViTA BPM64 (V1) is an automatic blood pressure monitor. Its blood pressure measurement technology has been proven to be accurate, with a 2-star Medaval rating. Blood pressure measurements are taken from the upper arm. It is intended for self-measurement and home use. This device has been discontinued by the manufacturer but may be available through certain outlets.
Assessment:
The technology used in the AViTA BPM64 (V1), to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval ★★ Recommendation Recent clinical validation; recent protocol
BP BIHS (UK and IRL) Self-measurement (Previous) Published evidence
BP Stride BP Self-measurement for adults (Preferred) Published evidence, but the device validated was the .
Device Family:
Validation Publications:

Kang YY, Chen Q, Liu CY, Li Y, Wang JG. Validation of the AVITA BPM64 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2018 Feb;23(1):45-8. Epub: 2017 Oct 18. doi: 10.1097/MBP.0000000000000296. PMID: 29049094.

ESH-IP:2010 - Pass General population

Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

AViTA BPM82

Menti A, Ntousopoulos V, Theodosiadi A, Stathopoulou P, Kollias A, Stergiou GS. Validation of the automated oscillometric upper-arm cuff home blood pressure monitor AVITA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard. Blood Press Monit. 2024 Apr 1;29(2):89-92. Epub: 2023 Nov 6. doi: 10.1097/MBP.0000000000000682. PMID: 37937605.

81060-2:2018/Amd 1:2020 - Pass General population

Menti A, Lempidaki P, Mariglis D, Antonogiannaki EC, Bakalakou KR, Komnianou A, Kollias A, Stergiou GS. Validation of the automated oscillometric home blood pressure monitor AViTA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Blood Press Monit. Epub: 2026 Feb 2. doi: 10.1097/MBP.0000000000000795. PMID: 41670419.

81060-2:2018/Amd 1:2020/Amd 2:2024 - Pass General population