🏠
📞  +353 1 6643768
 
📧  info@medaval.ie
A&D TM-2433
Device Model:
TM-2433
Manufacturer:
A&D Company Ltd., 3-23-14 Higashi-Ikebukuro, Toshima-Ku, 170-0013 Tokyo, JAPAN.
Measuring functions:
Blood pressure
Primary Client Use:
Intended for ambulatory blood pressure monitoring
Measurement Site:
Upper Arm
Measurement Occurrence:
Intermittent measurements at specified intervals or times
Availability:
Available Currently
Availability according to Countries or Regions:
Japan
Device Manual:
Description:
The A&D TM-2433 is an ambulatory blood pressure monitor. The accuracy of its blood pressure measurement technology has yet to be proven to MDR requirements. Blood pressure measurements are taken from the upper arm. It is intended for ambulatory blood pressure monitoring.
Assessment:
The technology used in the A&D TM-2433, to measure blood pressure, has been compared to a clinically validated device, but the results of the equivalence study have not been published and cannot be verified; nor can the protocol be checked for compliance with (EU) 2017/745 and MEDDEV 2.7/1 rev 4.
Recommendations:
Accuracy AssessmentRecommendationBasis
BP Medaval None Although equivalence to another device is claimed, this has not been tested to MDR requirements.
BP Japanese Society of Hypertension Self-measurement and professional use (2016-2022) Implied equivalence to the A&D TM-2430
Device Family:
A&D TM-2430J, A&D TM-2433J
Legend: J JHS Implied Equivalence
Validation Publications for Equivalent Devices:

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

A&D TM-2430

Palatini P, Frigo G, Bertolo O, Roman E, Da Cortà R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics. Blood Press Monit. 1998 Aug;3(4):255-60. PMID: 10212363.

BHS:1993 - Pass (A/A) General population

SP10:1992 - Pass General population

Yip GW, So HK, Li AM, Tomlinson B, Wong SN, Sung RY. Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol. Blood Press Monit. 2012 Apr;17(2):76-9. doi: 10.1097/MBP.0b013e328351d4a4. PMID: 22391521.

BHS:1993 - Pass (A/B) Children and adolescents (n=61)