Wellion Calla Dialog
    
          
          
        Device Name:
Calla Dialog
       Manufacturer:
MED TRUST Handelsgesellschaft m.b.H. (Wellion), Gewerbepark 10, A-7221 Marz, AUSTRIA.
       Measuring functions:
Blood glucose
       Primary Client Use:
Intended for self-measurement and home use
       Measurement Site:
Finger
       Measurement Occurrence:
Single measurements only
       Availability:
Available Currently
  Description:
The Wellion Calla Dialog is an automatic blood glucose meter. Medaval has not found evidence proving the accuracy of its blood glucose measurement technology. Blood glucose measurements are taken from the finger. It is intended for self-measurement and home use.
        Assessment:
The technology used in the Wellion Calla Dialog, to measure blood glucose, has failed in a clinical validation study, in a specific population, according to a recognised standard protocol, as published in a peer-reviewed publication.
        Recommendations:
        | Accuracy Assessment | Recommendation | Basis | |
| BG | Medaval | Not recommended | This device has failed a clinical validation. | 
Validation Publications:
Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: PMC3876330.
15197:2003 - Fail Using insufficient blood samples (n=31)