Device Name:GlucoRX Nexus
Device Model:TD-4277
Manufacturer:

Distributer: GlucoRx Ltd., 4 Wintonlea, Monument Way West, Woking, Surrey GU21 5EN, UNITED KINGDOM.

OEM: TaiDoc Technology Corporation, 6F - 127 Wugong 2nd Road, 24888 Wugu District, New Taipei City, TAIWAN PRC.

Measuring functions:Blood glucose
Primary Client Use:Intended for both professional use and self-measurement
Measurement Site:Finger and general Alternative Site Testing
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

The technology used in the TaiDoc GlucoRx Nexus (TD-4277), to measure blood glucose, is among the best rated for accuracy by Medaval Ltd. It has passed in clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★★ RecommendationRecent clinical validation; recent protocol; multiple clinical validations
MDR CriteriaSelf-measurement and professional usePublished evidence

Validation Publications

Salzsieder E, Berg S. Accuracy Evaluation of a CE-Marked System for Self-Monitoring of Blood Glucose With Three Reagent System Lots Following ISO 15197:2013. J Diabetes Sci Technol. 2016 Jan;10(1):238-9. Epub: 2015 Sep 22. doi: 10.1177/1932296815606471. PMID: 26394775. Available from: PMC4738220. No abstract available.

15197:2013 - Pass General population

TaiDoc GlucoRx Nexus (TD-4277), Menarini GlucoMen Nexus (TD-4277) and Aktivmed GlucoCheck XL (TD-4277) are described as being the same device.

Vanterpool GE. Accuracy of a Blood Glucose Monitoring System. J Diabetes Sci Technol. 2016 Jul;10(4):995-6. Epub: 2016 Feb 17. doi: 10.1177/1932296816630330. PMID: 26892507. Available from: PMC4928233. No abstract available.

15197:2013 - Pass General population