Roche Accu-Chek Mobile

Device Name:Accu-Chek Mobile
Manufacturer:Roche Diagnostics USA, 9115 Hague Road, PO Box 50457, Indianapolis, IN 46250-0457, UNITED STATES.
Measuring functions:Blood glucose
Primary Client Use:Intended for both professional use and self-measurement
Measurement Site:Finger and general Alternative Site Testing
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment: The technology used in the Roche Accu-Chek Mobile, to measure blood glucose, has passed in four validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.
Validation Publications

Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. PMID: 23063032. Available from: www.ncbi.nlm.nih.gov/pmc/articles/PMC3570840/pdf/dst-06-1060.pdf.

15197:2003 - Pass General population (Note: Maltose dependent test strip)

15197:2013 - Pass General population (Note: Maltose dependent test strip)

15197:2003 - Pass General population (Note: Maltose independent test strip)

15197:2013 - Pass General population (Note: Maltose independent test strip)

Pfützner A, Schipper C, Ramljak S, Flacke F, Sieber J, Forst T, Musholt PB. Evaluation of the effects of insufficient blood volume samples on the performance of blood glucose self-test meters. J Diabetes Sci Technol. 2013 Nov 1;7(6):1522-9. doi: 10.1177/193229681300700612. PMID: 24351178. Available from: www.ncbi.nlm.nih.gov/pmc/articles/PMC3876330/pdf/dst-07-1522.pdf.

15197:2003 - Pass Using insufficient blood samples (n=31)

Sølvik UØ, Risa M, Jacobsen CE, Monsen G, Sandberg S. Performance of 10 Systems for Self-Monitoring of Blood Glucose by Trained Healthcare Professionals and in the Hands of the Users. Clin Chem. 2015 May;61(5):772-4. Epub: 2015 Mar 3. doi: 10.1373/clinchem.2014.236760. PMID: 25737533. Available from: http://clinchem.aaccjnls.org/content/clinchem/61/5/772.full.pdf.

15197:2013 - Pass General population (Note: BLS-recorded measurements)

FDA:2016H - Pass General population (Note: BLS-recorded measurements (Failed 2013/99))

15197:2013 - Pass General population (Note: Subject-recorded measurements)

FDA:2016H - Pass General population (Note: Subject-recorded measurements)

15197:2003 - Pass General population

SKUP Evaluations 2009/74 and 2013/99