Home /  Omron HEM-9210T

Omron HEM-9210T

Device Model:HEM-9210T
Manufacturer:Sole: Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

The technology used in the Omron HEM-9210T, to measure blood pressure, is among the best rated for accuracy by Medaval Ltd. It has passed in clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★★ RecommendationRecent clinical validation; recent protocol; multiple clinical validations
BIHS (UK and IRL)NOT recommendedPublished evidence
ESH (Europe)Self-measurementPublished evidence
Stride BPSelf-measurement (Preferred)Published evidence
MDR CriteriaSelf-measurementPublished evidence

Validation Publications

Alpert BS. Validation of the Omron HEM-9210T by the ANSI/AAMI/ISO 81060-2:2013 with two novel cuffs: wide-range and extra-large. Blood Press Monit. 2017 Jun;22(3):166-8. Epub: 2017 Mar 1. doi: 10.1097/MBP.0000000000000250. PMID: 28252519.

81060-2:2013 - Pass General population

Takahashi H. 欧州高血圧学会国際プロトコル(ESH-IP)改訂2010年版による家庭血圧監視装置Omron EVOLV(HEM-7600T-E),HEM-9210T,およびM3Comfort(HEM-7143-E)の妥当性確認. [Validation of Home Blood Pressure-monitoring Devices Omron EVOLV (HEM-7600T-E), HEM-9210T, and M3 Comfort (HEM-7134-E) According to European Society of Hypertension International Protocol (ESH-IP) Revision 2010.] 日本臨床生理学会雑誌 [Jpn J Clin Physiol]. 2018;48(1):29-38.

ESH-IP:2010 - Fail General population

No abstract provided. "Fail" understood from BIHS rejection, which only considered this publication

Van Den Heuvel JFM, Lely AT, Franx A, Bekker MN. Validation of the iHealth Track and Omron HEM-9210T automated blood pressure devices for use in pregnancy. Pregnancy Hypertens. 2019 Jan. 37-41 p. doi: 10.1016/j.preghy.2018.10.008. PMID: 30825925.

ESH-IP:2010 - Pass Pregnancy (n=33) including pre-eclampsia (n=10)

Relevant Publications

Padwal RS. Validation of the Omron HEM-9210T by the ANSI/AAMI/ISO 81060-2:2013 with two novel cuffs: wide-range and extra-large. Blood Press Monit. 2017 Dec;22(6):379. doi: 10.1097/MBP.0000000000000295. PMID: 29112058.

The publication is a letter or correspondence in which the Omron HEM-9210T is referenced.

Alpert BS. Response to Regarding: validation of the Omron HEM-9210T by the ANSI/AAMI/ISO 81060-2 with two novel cuffs wide-range and extra-large Blood Press Monit. 2017 Dec;22(6):379-80. doi: 10.1097/MBP.0000000000000291. PMID: 29112059.

The publication is a letter or correspondence in which the Omron HEM-9210T is referenced.