Home /  Omron HEM-6320T

Omron HEM-6320T

Device Model:HEM-6320T
Manufacturer:Sole: Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600-8530, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Wrist
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

While the technology used in the Omron HEM-6320T, to measure blood pressure, has been claimed as being equivalent to that used in another device, not only has no evidence has been published to show that the devices were compared according to a protocol compliant with (EU) 2017/745 and MEDDEV 2.7/1 rev 4 but there is no evidence provided on a validation of the other device.

Accuracy AssessmentRecommendationBasis
MedavalNoneAlthough equivalence to another device is claimed, this has not been tested to MDR requirements.
Japanese Society of HypertensionSelf-measurement (2018)Implied equivalence to the Omron RS3 (HEM-6130-E), RS6 (HEM-6221-E) or RS8 (HEM-6310F-E)

Validation Publications for Equivalent Devices

Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.

Omron RS8 (HEM-6310F-E)

Takahashi H, Yoshika M, Yokoi T. Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010. Vasc Health Risk Manag. 2013;9:265-72. Epub: 2013 May 28. doi: 10.2147/VHRM.S44569. PMID: 23745050.

ESH-IP:2010 - Pass General population