Menarini GlucoMen Areo

Device Name:GlucoMen Areo
Manufacturer:A. Menarini Diagnostics S.r.l., Via Sette Santi 3, 50131 Firenze, ITALY.
Measuring functions:Blood glucose
Primary Client Use:Intended for both professional use and self-measurement
Measurement Site:Finger and general Alternative Site Testing
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment:

The technology used in the Menarini GlucoMen Areo, to measure blood glucose, is among the best rated for accuracy by Medaval Ltd. It has passed in clinical validation studies, between general and specific populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★★ RecommendationRecent clinical validation; recent protocol; multiple clinical validations
MDR CriteriaSelf-measurement and professional usePublished evidence

Validation Publications

Berti F, Scuffi C, Valgimigli F. Accuracy Evaluation of Two Blood Glucose Monitoring Systems Following ISO 15197:2013. J Diabetes Sci Technol. 2015 Nov;9(6):1354-6. Epub: 2015 Jul 29. doi: 10.1177/1932296815595986. PMID: 26224761. No abstract available.

15197:2013 - Pass General population

Jendrike N, Baumstark A, Pleus S, Liebing C, Kamecke U, Haug C, Freckmann G. Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. Curr Med Res Opin. Epub: 2018 Jul 19:1-11 p. doi: 10.1080/03007995.2018.1491832. PMID: 29927663.

15197:2013 - Pass General population

Tentolouris A, Eleftheriadou I, Grigoropoulou P, Nikoloudi M, Siami E, Tsilimigras DI, Tentolouris N. Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems Journal of Clinical and Diagnostic Research. 2018 Jun;12(6):OC10-OC14. doi: 10.7860/JCDR/2018/34055.11671. Available from: www.jcdr.net.

Ad Hoc protocol General population (Note: Tested in criteria resembling ISO 15197:2013.)