Fukuda Denshi DS-7000

Device Name:Dynascope
Device Model:DS-7000
Manufacturer:Fukuda Denshi, 17725-C NE 65th Street, Redmond, WA 98052, UNITED STATES.
Measuring functions:Blood pressure, Arrhythmias, SpO2, Temperature, ECG, Respiration
Primary Client Use:Intended for patient monitoring
Measurement Site:Upper Arm
Measurement Occurrence:Intermittent measurements at specified intervals or times
Availability:Available Currently
Assessment:

The technology used in the Fukuda Denshi DS-7000, to measure blood pressure, has passed in three clinical validation studies, in general populations, according to recognised standard protocols, as published in peer-reviewed publications.

Accuracy AssessmentRecommendationBasis
Medaval★★ RecommendationRecent clinical validation; older protocol; multiple clinical validations
ESH (Europe)Professional use in childrenPublished evidence

Validation Publications

Alpert BS, Blakely DW. Validation of the Fukuda Denshi DS-7000 patient monitor by AAMI standard testing. Blood Press Monit. 2009 Dec;14(6):274-6. doi: 10.1097/MBP.0b013e328332fd6d. PMID: 19923985.

BHS:1993 - Pass (A/A) In Normal Mode

BHS:1993 - Pass (A/A) In Quick Mode

SP10:2002 - Pass Adults and Children - In Normal Mode

SP10:2002 - Pass Adult and Children - In Quick Mode

SP10:2002 - Pass Infants (n=15) and Neonate (n=18)