A&D UA-855
The technology used in the A&D UA-855, to measure blood pressure, has passed in a clinical validation study, in a general population, according to a recognised standard protocol, as published in a peer-reviewed publication.
Accuracy Assessment | Recommendation | Basis | ||
---|---|---|---|---|
Medaval | Legacy approval | Older clinical validation; older protocol | ||
BIHS (UK and IRL) | Self-measurement | Published evidence | ||
ESH (Europe) | Self-measurement | Published evidence | ||
Japanese Society of Hypertension | Self-measurement (2016-2019) | Published evidence | ||
Stride BP | Self-measurement for adults | Published evidence (corrected from UA-85X) |
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.
ESH-IP:2002 - Pass General population
SP10:1987 - Pass General population
Note: This is a provisional list, as equivalence according to EU Regulation 2017/745 (e.g. MEDDEV 2.7/1 rev 4) is not proven. Accordingly, these publications are not used in the assessment of star-ratings.
A&D UA-851
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.
ESH-IP:2002 - Pass General population
SP10:1987 - Pass General population
A&D UA-852
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.
ESH-IP:2002 - Pass General population
SP10:1987 - Pass General population
A&D UA-853
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.
ESH-IP:2002 - Pass General population
SP10:1987 - Pass General population
A&D UA-854
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit. 2008 Dec;13(6):339-41. doi: 10.1097/MBP.0b013e32830d4b77. PMID: 19020424.
ESH-IP:2002 - Pass General population
SP10:1987 - Pass General population