Omron 705CP-II (HEM-759-E2)

Device Name:705CP-II
Device Model:HEM-759-E2
Manufacturer:Sole: Omron Corporation, Kyoto Head Office, Shiokoji Horikawa, Shimogyo ku, Kyoto 600 8530, JAPAN.
Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment: The technology used in the Omron 705CP-II (HEM-759-E2), to measure blood pressure, has passed in seven validation studies, between general and young populations, according to recognised standard protocols, as published in peer-reviewed publications.
Relevant Publications

Adler C, Ellert U, Neuhauser HK. Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values. Blood Press Monit. 2014 Apr;19(2):109-17. doi: 10.1097/MBP.0000000000000037. PMID: 24583967.

Measurements from the Omron 705CP-II (HEM-759-E2) are compared to those from the Datascope Accutorr Plus, though not in a formal validation.

Validation Publications for Equivalent Devices

Omron 705IT (HEM-759-E)

El Assaad MA, Topouchian JA, Asmar RG. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit. 2003 Jun;8(3):127-33. PMID: 12900590.

ESH-IP:2002 - Pass General population

Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: the Arsakion School Study. Blood Press Monit. 2006 Aug;11(4):229-34. PMID: 16810034.

ESH-IP:2002 - Pass Children amd adolescents (n=197)

SP10:2002 - Pass Children amd adolescents (n=197)

Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit. 2006 Feb;11(1):27-32. PMID: 16410738.

BHS:1993 - Pass (A/A) General population

SP10:2002 - Pass General population

Omron M6 (HEM-7001-E)

Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit. 2006 Jun;11(3):165-71. PMID: 16702826.

ESH-IP:2002 - Pass General population

Altunkan S, Ilman N, Kayatürk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults. Blood Press Monit. 2007 Aug;12(4):219-25. PMID: 17625394.

ESH-IP:2002 - Pass General population

Altunkan S, Iliman N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients. Blood Press Monit. 2008 Apr;13(2):117-22. doi: 10.1097/MBP.0b013e3282f3fb50. PMID: 18347447.

ESH-IP:2002 - Pass General population

Omron M6 (HEM-7211-E)

Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension. Vasc Health Risk Manag. 2011;7:709-17. Epub: 2011 Nov 30. doi: 10.2147/VHRM.S27193. PMID: 22174581. Available from: www.ncbi.nlm.nih.gov/pmc/articles/PMC3237100/pdf/vhrm-7-709.pdf.

ESH-IP:2010 - Pass General population