Microlife BP 3AS1-2

Device Model:BP 3AS1-2
Manufacturer:Sole: Microlife Corporation, 9F, 431 RuiGuang Road, NeiHu, Taipei 11492, TAIWAN PRC.
Measuring functions:Blood pressure
Primary Client Use:Intended for self-measurement and home use
Measurement Site:Upper Arm
Measurement Occurrence:Single measurements only
Availability:Available Currently
Assessment: The technology used in the Microlife BP 3AS1-2, to measure blood pressure, has passed in three validation studies, between general and pregnancy populations, according to recognised standard protocols, as published in peer-reviewed publications.
Validation Publications

de Greeff A, Nathan H, Stafford N, Liu B, Shennan AH. Development of an accurate oscillometric blood pressure device for low resource settings. Blood Press Monit. 2008 Dec;13(6):342-8. doi: 10.1097/MBP.0b013e32830fd07c. PMID: 19020425.

ESH-IP:2002 - Pass General population

Nathan HL, de Greeff A, Hezelgrave NL, Chappell LC, Shennan AH. An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2. Blood Press Monit. 2015 Feb;20(1):52-5. doi: 10.1097/MBP.0000000000000086. PMID: 25243711.

BHS:1993 - Pass (B/A) Pregnancy and pre-eclampsia (n=45)

81060-2:2013 - Pass Pregnancy and pre-eclampsia (n=45)

Nathan HL, de Greeff A, Hezelgrave NL, Chappell LC, Shennan AH. Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure. Blood Press Monit. 2015 Oct;20(5):299-302. Epub: 2015 May 5. doi: 10.1097/MBP.0000000000000134. PMID: 26020367.

BHS:1993 - Pass (A/A) Low BP pregnancy (n=30)

81060-2:2013 - Pass Low BP pregnancy (n=30)